Industrial Scientific Australia Pty Ltd—GasBadge Plus version 2.2 and earlier instruments

PRA number

2012/13003

Published date

Product description

Single-gas portable gas detection instrument
Identifying features
The following part numbers are covered by this recall: GasBadge® Plus – Carbon Monoxide 18100050‐1, 18100050‐1xx, IPLU501 GasBadge® Plus – Hydrogen Sulfide 18100050‐2, 18100050‐2xx, IPLU502 GasBadge® Plus – Oxygen 18100050‐3, 18100050‐3xx, IPLU503 GasBadge® Plus – Nitrogen Dioxide 18100050‐4, 18100050‐4xx, IPLU504 GasBadge® Plus – Sulfur Dioxide 18100050‐5, 18100050‐5xx, IPLU505 xx = optional alphanumeric designation

Other identifying numbers

The following part numbers are covered by this recall:
GasBadge® Plus – Carbon Monoxide 18100050‐1, 18100050‐1xx, IPLU501
GasBadge® Plus – Hydrogen Sulfide 18100050‐2, 18100050‐2xx, IPLU502
GasBadge® Plus – Oxygen 18100050‐3, 18100050‐3xx, IPLU503
GasBadge® Plus – Nitrogen Dioxide 18100050‐4, 18100050‐4xx, IPLU504
GasBadge® Plus – Sulfur Dioxide 18100050‐5, 18100050‐5xx, IPLU505
xx = optional alphanumeric designation

Reasons the product is recalled

The GasBadge Plus single-gas instrument may not properly indicate a low battery condition, resulting in a failure to "alarm" or "instrument shut down". This condition is exacerbated when the instrument is used in temperatures below room temperature.

The hazards to consumers

The defect may cause the user to be exposed to noxious gases without being alerted, resulting in poisoning.

What consumers should do

Customers should cease using the product and contact their local dealer to organise for a replacement unit. Industrial Scientific is also offering customers a free upgrade to GasBadge® Pro single-gas instruments.

Customers can contact Industrial Scientific on 03 9644 7777 (Australia and New Zealand) or email gbplusrecall@indsci.com

Traders who sold this product

Airmet Scientific Pty Ltd and Protector Alsafe Pty Ltd

Where the product was sold

Nationally

Dates available for sale

-

Responsible regulator

Australian Competition and Consumer Commission is the responsible regulator for this recall.