Fluke—37X Series Clamp Meters

PRA number

2013/13457

Published date

Product description

Fluke 37X Series Clamp Meters
Fluke 373 Clamp Meter
Fluke 374 Clamp Meter
Fluke Clamp Meter
Identifying features
Models affected: Fluke-376 Fluke-375 Fluke-374 Fluke-373 Please see the attached list for further information on manufacture dates and serial numbers.

Other identifying numbers

Models affected:
Fluke-376
Fluke-375
Fluke-374
Fluke-373
Please see the attached list for further information on manufacture dates and serial numbers.

Reasons the product is recalled

A screw connecting the printed circuit assembly to the test lead input jack can work itself loose and result in intermittent voltage readings, including a ‘0’ reading when voltage is present.

The hazards to consumers

If the clamp meter is not reading correctly, users may be exposed to a potential electrocution hazard.

What consumers should do

1. Stop using your affected 37X series clamp meter(s) even if you have not experienced problems.

2. Verify that your clamp meter is affected using the serial numbers provided in the table.

3. If the product falls within the serial number range, return the unit to Celemetrix for repair.