IBD Medical — Dario Blood Glucose Tracker App Android App Versions 4.3.0-4.3.2

PRA number

2019/17815

Published date

Product description

Dario Blood Glucose Tracker App
Android App Versions 4.3.0-4.3.2

ATRG 229859
(IBD Medical - Software IVDs)

An in vitro diagnostic medical device (IVD)
Identifying features

Other identifying numbers

Android App Versions 4.3.0-4.3.2

Reasons the product is recalled

Potential issue with Android App versions of Dario Blood Glucose Tracker (4.3.0-4.3.2), regarding duplicate logging of a blood glucose level reading. The issue may cause blood glucose level readings that are transferred from the measurement screen to the 'Logbook' being replaced by an older blood glucose measurement and being displayed as a duplicate reading.

The hazards to consumers

If inaccurate blood glucose level reading from the 'Data Entry' or 'Logbook' screens are used to determine the insulin dose or make treatment decisions, this may lead to incorrect dosing decisions, which can lead to serious health complications.

What consumers should do

Update the latest Dario - Blood Glucose Tracker Android App version as follows:
Open the Google Play store
Search for 'Dario Blood Glucose Tracker' App
Press the 'Uninstall' button
Press 'OK' to confirm the question "Do you want to uninstall this app?"
Once completed, press 'Install' and verify 4.3.3 version (or above) has been installed by pressing the 'About this App' link.

For further information, contact IBD Medical on 02 8398 7520

Supplier

Traders who sold this product

Google Play Store

Where the product was sold

Nationally
Online

Responsible regulator

Therapeutic Goods Administration is the responsible regulator for this recall.