Fluke Australia Pty Ltd—Fluke 33X Clamp Meters

PRA number

2009/10778

Published date

Product description

Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters manufactured between 3 March 2008 and 27 February 2009.
Identifying features
Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters

Other identifying numbers

Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters

Reasons the product is recalled

A defect has the potential, over time, to cause the meter to display a low or no voltage reading on a circuit energised with a hazardous voltage. In this situation it is imperative that the practice of testing the meter on a known live circuit before and after checking the circuit under test is maintained. Should the meter not pass the "known live circuit test" its use should be discontinued immediately, identified as faulty and removed from service.

The hazards to consumers

Electrocution due to incorrect reading

What consumers should do

Cease using and immediately return suspect products to the place of purchase to be replaced free of charge. For further information visit www.fluke.com.au or phone (02) 8850 3333.

Where the product was sold

Nationally

Recall advertisements and supporting documentation

Responsible regulator

Australian Competition and Consumer Commission is the responsible regulator for this recall.