Alphacare Pty Ltd — Bock Nursing Care Beds - Battery

PRA number

2020/18174

Published date

Product description

Bock Nursing Care Beds - Battery
Model Number: 263
Serial Number: 91391986-000

ARTG 209294
(Alphacare Pty Ltd - Bed, general-purpose)

Reasons the product is recalled

In beds with double drives, manufactured in the period from January 2012 to May 2013, some batteries may contain diodes which may be defective.

The hazards to consumers

Failure of the diode may lead to damage of the emergency lowering battery. If the bed cannot be lowered this may lead to injury.

What consumers should do

Consumers with impacted stock should contact Alphacare to organise correction. Alphacare will arrange a technician to visit and make the necessary corrections.

For further information, contact Alphacare Customer Service on 1300 930 930 or via the Alphacare website at https://alphacare.com.au/contact/

Where the product was sold

New South Wales

Dates available for sale

-

Responsible regulator

Therapeutic Goods Administration is the responsible regulator for this recall.

Product category