Abbott Australasia Pty Ltd—ClearStar Enteral Pump

PRA number

2007/9363

Published date

Product description

List number M771. Serial number below M2701564. ARTG number 71877.

Reasons the product is recalled

Since 2000 there have been 12 complaints (worldwide) reporting under delivery for these pumps; the pumps were confirmed to have cracked or broken adaptor brackets. There was one reported adverse event (in the US) involving weight loss over a 3 month period.

The hazards to consumers

Illness or mistreatment (Class II).

What consumers should do

Please contact Abbott Australasia Pty Ltd Customer Service telephone number 1800 225 311.

Supplier

Abbott Australasia Pty Ltd

Where the product was sold

Nationally

Responsible regulator

Therapeutic Goods Administration is the responsible regulator for this recall.